Niscomed M1000 Multipara Patient Monitor Firmware

4 CVEs affecting Niscomed M1000 Multipara Patient Monitor Firmware. Latest disclosed: 2020-08-26. Critical: 0, High: 2.

Top CVEs affecting Niscomed M1000 Multipara Patient Monitor Firmware
CVESeverityScorePublishedSummary
CVE-2020-15482High7.82020-08-26An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the…
CVE-2020-15484High7.52020-08-26An issue was discovered on Nescomed Multipara Monitor M1000 devices. The internal storage of the underlying Linux system stores data in cleartext, without inte…
CVE-2020-15483Medium6.82020-08-26An issue was discovered on Nescomed Multipara Monitor M1000 devices. The physical UART debug port provides a shell, without requiring a password, with complete…
CVE-2020-15485Medium5.52020-08-26An issue was discovered on Nescomed Multipara Monitor M1000 devices. The onboard Flash memory stores data in cleartext, without integrity protection against ta…