Auth bypass in Niscomed M1000 Multipara Patient Monitor Firmware
CVE-2020-15482
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the admin account. This allows an attacker to gain root access to the device over the local network.
Vulnerability class: Broken Authentication
Published · last modified .
CVSS v3 metric
CVSS v3 base score 7.8 (High). Vector: CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS 0.002, 10.0th percentile (last changed ); unchanged over 30 days. About EPSS.
Affected products
Weakness classification (CWE)
References
- www.niscomed.com/multipara-monitor.html (Product)
- payatu.com/advisory/unauthenticated-telnet-service-in-niscomed-patient-monitor (Third Party Advisory)