Auth bypass in Niscomed M1000 Multipara Patient Monitor Firmware
CVE-2020-15483
- CVSS v3
- 6.8 Medium
- Public exploits
- None known
- EPSS
- 0.004 · 37.2th percentile
- Published
- 2020-08-26
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The physical UART debug port provides a shell, without requiring a password, with complete access.
Vulnerability class: Broken Authentication
Published · last modified .
CVSS v3 metric
CVSS v3 base score 6.8 (Medium). Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS 0.004, 37.2th percentile (last changed ); unchanged over 30 days. About EPSS.
Affected products
Weakness classification (CWE)
References
- www.niscomed.com/multipara-monitor.html (Product)
- payatu.com/advisory/unauthenticated-uart-root-shell--in-niscomed-patient-monitor (Exploit, Third Party Advisory, Broken Link)