Niscomed M1000 Multipara Patient Monitor
4 CVEs affecting Niscomed M1000 Multipara Patient Monitor. Latest disclosed: 2020-08-26. Critical: 0, High: 2.
| CVE | Severity | Score | Published | Summary |
|---|---|---|---|---|
CVE-2020-15482 | High | 7.8 | 2020-08-26 | An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the… |
CVE-2020-15484 | High | 7.5 | 2020-08-26 | An issue was discovered on Nescomed Multipara Monitor M1000 devices. The internal storage of the underlying Linux system stores data in cleartext, without inte… |
CVE-2020-15483 | Medium | 6.8 | 2020-08-26 | An issue was discovered on Nescomed Multipara Monitor M1000 devices. The physical UART debug port provides a shell, without requiring a password, with complete… |
CVE-2020-15485 | Medium | 5.5 | 2020-08-26 | An issue was discovered on Nescomed Multipara Monitor M1000 devices. The onboard Flash memory stores data in cleartext, without integrity protection against ta… |