Baxter Sigma_spectrum_infusion_system
10 CVEs affecting Baxter Sigma_spectrum_infusion_system. Latest disclosed: 2020-06-29. Critical: 8, High: 0.
| CVE | Severity | Score | Published | Summary |
|---|---|---|---|---|
CVE-2020-12047 | Critical | 9.8 | 2020-06-29 | The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless config… |
CVE-2020-12045 | Critical | 9.8 | 2020-06-29 | The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet ser… |
CVE-2020-12043 | Critical | 9.8 | 2020-06-29 | The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operatio… |
CVE-2020-12040 | Critical | 9.8 | 2020-06-29 | Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an un… |
CVE-2014-5433 | Critical | 9.8 | 2019-03-26 | An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum I… |
CVE-2014-5432 | Critical | 9.8 | 2019-03-26 | Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH withou… |
CVE-2014-5434 | Critical | 9.8 | 2019-03-26 | Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credent… |
CVE-2020-12041 | Critical | 9.4 | 2020-06-29 | The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits… |
CVE-2014-5431 | Medium | 6.8 | 2019-03-26 | Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provide… |
CVE-2020-12039 | Low | 2.4 | 2020-06-29 | Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hard… |