Vulnerability in Baxter Sigma Spectrum Infusion System Firmware
CVE-2020-12040
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 9.8 (Critical). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS 0.010, 62.5th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Baxter Sigma Spectrum Infusion System Firmware — versions 8.0
- Baxter Sigma Spectrum Infusion Pumps — versions Sigma Spectrum v6.x model 35700BAX, Baxter Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module v9,11,13,14,15,16,v20D29,v20D30,v20D31,v22D24, Baxter Spectrum v8.x with Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum LVP v8.x w/Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24
Weakness classification (CWE)
References
- www.us-cert.gov/ics/advisories/icsma-20-170-04 (US Government Resource, Third Party Advisory)