Vulnerability in Baxter Sigma Spectrum Infusion System Firmware
CVE-2020-12039
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 2.4 (Low). Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N.
EPSS exploit prediction
EPSS 0.003, 28.8th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Baxter Sigma Spectrum Infusion System Firmware — versions 8.0
- Baxter Sigma Spectrum Infusion Pumps — versions Sigma Spectrum v6.x model 35700BAX,Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module (WBM) v9,v11,v13,v14,v15,v16,v20D29,v20D30,v20D31,v22D24, Spectrum v8.x w/WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum LVP v8.x with WBM v17, v20D29,v20D30,v20D31,and v22D24
Weakness classification (CWE)
References
- www.us-cert.gov/ics/advisories/icsma-20-170-04 (US Government Resource, Third Party Advisory)