Vulnerability in Eppendorf Bioflo 320

CVE-2026-7251

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 9.8 (Critical). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.

EPSS exploit prediction

EPSS: 0.007 (52.8th percentile), scored .

Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 53rd percentile — 52.8% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.

EPSS trend (30 days)EPSS over the last 30 days for CVE-2026-7251: held from 0.005 to 0.007.

EPSS over last 30 days · oldest: 0.005 · newest: 0.007 · change: +0.002

Affected products

Weakness classification (CWE)

References

Frequently asked questions

What is CVE-2026-7251?
CVE-2026-7251 is a critical-severity vulnerability in Eppendorf Bioflo 320, classified under Use of Hard-coded Password. CVSS score: 9.8/10. Published 2026-05-26.
How severe is CVE-2026-7251?
Critical severity. CVSS v3 base score is 9.8 out of 10.