Auth bypass in Quanovate Tech Inc. (Operating As Mira / Care) Android App
CVE-2026-66875
- CVSS v3
- 8.8 High
- Public exploits
- None known
- EPSS
- 0.004 · 35.3th percentile
- Published
- 2026-08-11
In the Mira hormone monitor device firmware v1.7.1.47 build 01070147, a remote unauthenticated attacker within BLE range (approximately 10–30 meters) can silently rebind the device to an attacker-controlled account, extract stored hormone measurements in cleartext, cause a denial-of-service via malformed or undocumented command opcodes, and passively track the user via a static random BLE address that never rotates.
Vulnerability class: Broken Authentication
Published · last modified .
CVSS v3 metric
CVSS v3 base score 8.8 (High). Vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
- Attack Vector
- Adjacent
AV:A - Attack Complexity
- Low
AC:L - Privileges Required
- None
PR:N - User Interaction
- None
UI:N - Scope
- Unchanged
S:U - Confidentiality
- High
C:H - Integrity
- High
I:H - Availability
- High
A:H
EPSS exploit prediction
EPSS 0.004, 35.3th percentile (last changed ); unchanged over 30 days. About EPSS.
Affected products
- Quanovate Tech Inc. (Operating As Mira / Care) Android App — versions 4.5.15.4, 4.5.18
- Quanovate Tech Inc. (Operating As Mira / Care) Firmware — versions 1.7.1.47, 01.07.01.53