Vulnerability in Philips Intellispace Cardiovascular (Iscv)
CVE-2025-2229
A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 7.7 (High). Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N.
EPSS exploit prediction
EPSS: 0.001 (1.9th percentile), scored .
Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 2nd percentile — 1.9% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.
EPSS over last 30 days · oldest: 0.001 · newest: 0.001 · change: 0.000
Affected products
- Philips Intellispace Cardiovascular (Iscv) — versions 0
Weakness classification (CWE)
References
Frequently asked questions
- What is CVE-2025-2229?
- CVE-2025-2229 is a high-severity vulnerability in Philips Intellispace Cardiovascular (Iscv), classified under CWE-1391. CVSS score: 7.7/10. Published 2025-03-13.
- How severe is CVE-2025-2229?
- High severity. CVSS v3 base score is 7.7 out of 10.