Vulnerability in Radiometer Medical Aps Abl800 Basic And Flex Analyzers
CVE-2025-14096
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
EPSS: 0.001 (2.0th percentile) — read the EPSS interpretation.
CVSS v3 metric
CVSS v3 base score 8.4 (High). Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
Affected products
- Radiometer Medical Aps Abl800 Basic And Flex Analyzers — versions Windows 7 Operating system, Windows XP Operating system
- Radiometer Medical Aps Abl90 Flex And Plus Analyzers — versions Windows 7 Operating system, Windows XP Operating system
- Radiometer Medical Aps Aqt90 Flex Analyzers — versions Windows 7 Operating system, Windows XP Operating system
Weakness classification (CWE)
References
Frequently asked questions
- What is CVE-2025-14096?
- CVE-2025-14096 is a high-severity vulnerability in Radiometer Medical Aps Abl800 Basic And Flex Analyzers, classified under CWE-250. CVSS score: 8.4/10. Published 2025-12-17.
- How severe is CVE-2025-14096?
- High severity. CVSS v3 base score is 8.4 out of 10.