Vulnerability in Radiometer Medical Aps Abl800 Basic And Flex Analyzers

CVE-2025-14096

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.

Other related CVE's are CVE-2025-14095 & CVE-2025-14097.

Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required Configuration for Exposure:

Attacker requires physical access to the analyzer.

Temporary work Around:
Only authorized people can physically access the analyzer.

Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status:

Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 8.4 (High). Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.

EPSS exploit prediction

EPSS 0.001, 3.0th percentile (last changed ); unchanged over 30 days. About EPSS.

Affected products

Weakness classification (CWE)

References