Vulnerability in Radiometer Medical Aps Abl800 Basic And Flex Analyzers

CVE-2025-14096

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

EPSS: 0.001 (2.0th percentile) — read the EPSS interpretation.

CVSS v3 metric

CVSS v3 base score 8.4 (High). Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.

Affected products

Weakness classification (CWE)

References

Frequently asked questions

What is CVE-2025-14096?
CVE-2025-14096 is a high-severity vulnerability in Radiometer Medical Aps Abl800 Basic And Flex Analyzers, classified under CWE-250. CVSS score: 8.4/10. Published 2025-12-17.
How severe is CVE-2025-14096?
High severity. CVSS v3 base score is 8.4 out of 10.