Vulnerability in Radiometer Medical Aps Abl800 Basic And Flex Analyzers

CVE-2025-14095

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.  Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 6.8 (Medium). Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.

EPSS exploit prediction

EPSS: 0.002 (5.3th percentile), scored .

Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 5th percentile — 5.3% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.

EPSS trend (30 days)EPSS over the last 30 days for CVE-2025-14095: held from 0.002 to 0.002.

EPSS over last 30 days · oldest: 0.002 · newest: 0.002 · change: +0.000

Affected products

Weakness classification (CWE)

References

Frequently asked questions

What is CVE-2025-14095?
CVE-2025-14095 is a medium-severity vulnerability in Radiometer Medical Aps Abl800 Basic And Flex Analyzers, classified under Improper Access Control. CVSS score: 6.8/10. Published 2025-12-17.
How severe is CVE-2025-14095?
Medium severity. CVSS v3 base score is 6.8 out of 10.