Vulnerability in Contec Health Cms8000 Patient Monitor
CVE-2025-1204
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Published · last modified .
EPSS exploit prediction
EPSS: 0.005 (39.2th percentile), scored .
Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 39th percentile — 39.2% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.
EPSS over last 30 days · oldest: 0.005 · newest: 0.005 · change: +0.000
Affected products
- Contec Health Cms8000 Patient Monitor — versions 0
Weakness classification (CWE)
References
- claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a… (third-party-advisory)
- www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01 (government-resource)