Vulnerability in Contec Health Cms8000 Patient Monitor

CVE-2025-0683

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text
patient data to a hard-coded public IP address when a patient is hooked
up to the monitor. This could lead to a leakage of confidential patient
data to any device with that IP address or an attacker in a
machine-in-the-middle scenario.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 5.9 (Medium). Vector: CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N.

EPSS exploit prediction

EPSS 0.008, 56.4th percentile (last changed ); unchanged over 30 days. About EPSS.

Affected products

Weakness classification (CWE)

References