Vulnerability in Contec Health Cms8000 Patient Monitor
CVE-2025-0626
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 7.5 (High). Vector: CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS: 0.011 (65.6th percentile), scored .
Low probability of exploitation in the next 30 days, but still tracked. 66th percentile — 65.6% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.
EPSS over last 30 days · oldest: 0.012 · newest: 0.011 · change: -0.000
Affected products
- Contec Health Cms8000 Patient Monitor — versions All versions
Weakness classification (CWE)
References
Frequently asked questions
- What is CVE-2025-0626?
- CVE-2025-0626 is a high-severity vulnerability in Contec Health Cms8000 Patient Monitor, classified under Hidden Functionality. CVSS score: 7.5/10. Published 2025-01-30.
- How severe is CVE-2025-0626?
- High severity. CVSS v3 base score is 7.5 out of 10.