Vulnerability in Baxter Life2000 Ventilation System
CVE-2024-48970
The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 9.3 (Critical). Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H.
EPSS exploit prediction
EPSS: 0.002 (10.4th percentile), scored .
Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 10th percentile — 10.4% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.
EPSS over last 30 days · oldest: 0.002 · newest: 0.002 · change: 0.000
Affected products
- Baxter Life2000 Ventilation System — versions 06.08.00.00 and prior
Weakness classification (CWE)
References
Frequently asked questions
- What is CVE-2024-48970?
- CVE-2024-48970 is a critical-severity vulnerability in Baxter Life2000 Ventilation System, classified under CWE-1191. CVSS score: 9.3/10. Published 2024-11-14.
- How severe is CVE-2024-48970?
- Critical severity. CVSS v3 base score is 9.3 out of 10.