Auth bypass in Medtronic Insterstim Applications

CVE-2023-25931

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 6.4 (Medium). Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H.

EPSS exploit prediction

EPSS 0.003, 18.9th percentile (scored ); unchanged over 30 days. About EPSS.

Affected products

Weakness classification (CWE)

References