RCE in B. Braun Space Battery Pack Sp With Wi-Fi

CVE-2023-0888

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Published · last modified .

CVSS v3 metric

CVSS v3 base score 4.9 (Medium). Vector: CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N.

EPSS exploit prediction

EPSS: 0.006 (47.2th percentile), scored .

Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 47th percentile — 47.2% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.

EPSS trend (30 days)EPSS over the last 30 days for CVE-2023-0888: held from 0.006 to 0.006.

EPSS over last 30 days · oldest: 0.006 · newest: 0.006 · change: 0.000

Affected products

Weakness classification (CWE)

References

Frequently asked questions

What is CVE-2023-0888?
CVE-2023-0888 is a medium-severity vulnerability in B. Braun Space Battery Pack Sp With Wi-Fi, classified under Eval Injection. CVSS score: 4.9/10. Published 2023-03-13.
How severe is CVE-2023-0888?
Medium severity. CVSS v3 base score is 4.9 out of 10.