Vulnerability in Contec Health Cms8000 Icu Ccu Vital Signs Patient Monitor
CVE-2022-36385
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 6.8 (Medium). Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS 0.004, 36.2th percentile (last changed ); unchanged over 30 days. About EPSS.
Affected products
Weakness classification (CWE)
References
- www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 (US Government Resource, Third Party Advisory, Mitigation)