Deserialization in Siemens Biograph Horizon Pet/ct Systems
CVE-2022-29875
A vulnerability has been identified in Biograph Horizon PET/CT Systems (All VJ30 versions < VJ30C-UD01), MAGNETOM Family (NUMARIS X: VA12M, VA12S, VA10B, VA20A, VA30A, VA31A), MAMMOMAT Revelation (All VC20 versions < VC20D), NAEOTOM Alpha (All VA40 versions < VA40 SP2), SOMATOM X.cite (All versions < VA30 SP5 or VA40 SP2), SOMATOM X.creed (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.All (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Now (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Open Pro (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Sim (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Top (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Up (All versions < VA30 SP5 or VA40 SP2), Symbia E/S (All VB22 versions < VB22A-UD03), Symbia Evo (All VB22 versions < VB22A-UD03), Symbia Intevo (All VB22 versions < VB22A-UD03), Symbia T (All VB22 versions < VB22A-UD03), Symbia.net (All VB22 versions < VB22A-UD03), syngo.via VB10 (All versions), syngo.via VB20 (All versions), syngo.via VB30 (All versions), syngo.via VB40 (All versions < VB40B HF06), syngo.via VB50 (All versions), syngo.via VB60 (All versions < VB60B HF02). The application deserialises untrusted data without sufficient validations that could result in an arbitrary deserialization. This could allow an unauthenticated attacker to execute code in the affected system if ports 32912/tcp or 32914/tcp are reachable.
Vulnerability class: Insecure Deserialization
EPSS: 0.017 (74.9th percentile) — read the EPSS interpretation.
CVSS v3 metric
CVSS v3 base score 9.8 (Critical). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
Affected products
- Siemens Biograph Horizon Pet/ct Systems — versions All VJ30 versions < VJ30C-UD01
- Siemens Biograph_horizon_pet\/ct_systems
- Siemens Biograph_horizon_pet\/ct_systems_firmware
- Siemens Magnetom Family — versions NUMARIS X: VA12M, VA12S, VA10B, VA20A, VA30A, VA31A
- Siemens Magnetom_numaris_x
- Siemens Magnetom_numaris_x_firmware — versions va10b, va12m, va12s
- Siemens Mammomat Revelation — versions All VC20 versions < VC20D
- Siemens Mammomat_revelation
- Siemens Mammomat_revelation_firmware
- Siemens Naeotom Alpha — versions All VA40 versions < VA40 SP2
Weakness classification (CWE)
References
- productcert@siemens.com (x_refsource_MISC, Mitigation, Vendor Advisory)
Frequently asked questions
- What is CVE-2022-29875?
- CVE-2022-29875 is a critical-severity vulnerability in Siemens Biograph Horizon Pet/ct Systems, classified under Deserialization of Untrusted Data. CVSS score: 9.8/10. Published 2022-06-01.
- How severe is CVE-2022-29875?
- Critical severity. CVSS v3 base score is 9.8 out of 10.