Deserialization in Siemens Biograph Horizon Pet/ct Systems
CVE-2022-29875
- CVSS v3
- 9.8 Critical
- Public exploits
- None known
- EPSS
- 0.019 · 79.9th percentile
- Published
- 2022-06-01
A vulnerability has been identified in Biograph Horizon PET/CT Systems (All VJ30 versions < VJ30C-UD01), MAGNETOM Family (NUMARIS X: VA12M, VA12S, VA10B, VA20A, VA30A, VA31A), MAMMOMAT Revelation (All VC20 versions < VC20D), NAEOTOM Alpha (All VA40 versions < VA40 SP2), SOMATOM X.cite (All versions < VA30 SP5 or VA40 SP2), SOMATOM X.creed (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.All (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Now (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Open Pro (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Sim (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Top (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Up (All versions < VA30 SP5 or VA40 SP2), Symbia E/S (All VB22 versions < VB22A-UD03), Symbia Evo (All VB22 versions < VB22A-UD03), Symbia Intevo (All VB22 versions < VB22A-UD03), Symbia T (All VB22 versions < VB22A-UD03), Symbia.net (All VB22 versions < VB22A-UD03), syngo.via VB10 (All versions), syngo.via VB20 (All versions), syngo.via VB30 (All versions), syngo.via VB40 (All versions < VB40B HF06), syngo.via VB50 (All versions), syngo.via VB60 (All versions < VB60B HF02). The application deserialises untrusted data without sufficient validations that could result in an arbitrary deserialization. This could allow an unauthenticated attacker to execute code in the affected system if ports 32912/tcp or 32914/tcp are reachable.
Vulnerability class: Insecure Deserialization
Published · last modified .
CVSS v3 metric
CVSS v3 base score 9.8 (Critical). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
- Attack Vector
- Network
AV:N - Attack Complexity
- Low
AC:L - Privileges Required
- None
PR:N - User Interaction
- None
UI:N - Scope
- Unchanged
S:U - Confidentiality
- High
C:H - Integrity
- High
I:H - Availability
- High
A:H
EPSS exploit prediction
EPSS 0.019, 79.9th percentile (last changed ); unchanged over 30 days. About EPSS.
Affected products
- Siemens Biograph Horizon Pet/ct Systems — versions All VJ30 versions < VJ30C-UD01
- Siemens Biograph Horizon Pet\/ct Systems Firmware
- Siemens Magnetom Family — versions NUMARIS X: VA12M, VA12S, VA10B, VA20A, VA30A, VA31A
- Siemens Magnetom Numaris X Firmware — versions va10b, va12m, va12s
- Siemens Mammomat Revelation — versions All VC20 versions < VC20D
- Siemens Mammomat Revelation Firmware
- Siemens Naeotom Alpha — versions All VA40 versions < VA40 SP2
- Siemens Naeotom Alpha Firmware — versions va40
- Siemens Somatom Go.all — versions All versions < VA30 SP5 or VA40 SP2
- Siemens Somatom Go.all Firmware — versions va30, va40
Weakness classification (CWE)
References
- www.siemens-healthineers.com/support-documentation/cybersecurity/shsa-455016 (Mitigation, Vendor Advisory)