Auth bypass in Illumina Iseq_100
CVE-2022-1521
LRM does not implement authentication or authorization by default. A malicious actor can inject, replay, modify, and/or intercept sensitive data.
EPSS: 0.010 (59.2th percentile) — read the EPSS interpretation.
CVSS v3 metric
CVSS v3 base score 9.1 (Critical). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N.
Affected products
- Illumina Iseq_100
- Illumina Iseq 100 Instrument — versions LRM Versions 1.3 to 3.1
- Illumina Local_run_manager
- Illumina Miniseq
- Illumina Miniseq Instrument — versions LRM Versions 1.3 to 3.1
- Illumina Miseq
- Illumina Miseq Dx — versions LRM Versions 1.3 to 3.1
- Illumina Miseq_dx
- Illumina Miseq Instrument — versions LRM Versions 1.3 to 3.1
- Illumina Nextseq_500
Weakness classification (CWE)
References
- ics-cert@hq.dhs.gov (US Government Resource, Third Party Advisory, x_refsource_MISC)
Frequently asked questions
- What is CVE-2022-1521?
- CVE-2022-1521 is a critical-severity vulnerability in Illumina Iseq_100, classified under Improper Access Control. CVSS score: 9.1/10. Published 2022-06-24.
- How severe is CVE-2022-1521?
- Critical severity. CVSS v3 base score is 9.1 out of 10.