Vulnerability in Bbraun Infusomat Large Volume Pump 871305u
CVE-2021-33886
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 8.1 (High). Vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N.
EPSS exploit prediction
EPSS: 0.008 (55.8th percentile), scored .
Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 56th percentile — 55.8% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.
EPSS over last 30 days · oldest: 0.008 · newest: 0.008 · change: 0.000
Affected products
Weakness classification (CWE)
References
- www.bbraunusa.com/en.htm (Broken Link)
- www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-unc… (Exploit, Third Party Advisory)
Frequently asked questions
- What is CVE-2021-33886?
- CVE-2021-33886 is a high-severity vulnerability in Bbraun Infusomat Large Volume Pump 871305u, classified under Use of Externally-Controlled Format String. CVSS score: 8.1/10. Published 2021-08-25.
- How severe is CVE-2021-33886?
- High severity. CVSS v3 base score is 8.1 out of 10.