Vulnerability in Bbraun Spacecom2

CVE-2021-33885

An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 10.0 (Critical). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:N.

EPSS exploit prediction

EPSS 0.056, 93.0th percentile (scored ); unchanged over 30 days. About EPSS.

Affected products

Weakness classification (CWE)

References