The standard access path of the IntelliBridge EC 40 and 60 Hub (C.00.04 and prior) requires authentication, but the product has an alternate path or channel that does not require authentication.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 8.1 (High). Vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H.
EPSS exploit prediction
EPSS 0.005, 42.2th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Philips Intellibridge Ec40 Firmware
- Philips Intellibridge Ec 40 Hub — versions unspecified
- Philips Intellibridge Ec80 Firmware
- Philips Intellibridge Ec 80 Hub — versions unspecified
Weakness classification (CWE)
References
- www.cisa.gov/uscert/ics/advisories/icsma-21-322-01 (US Government Resource, Third Party Advisory)