Vulnerability in Fresenius Kabi Agilia Link+
CVE-2021-23233
Sensitive endpoints in Fresenius Kabi Agilia Link+ v3.0 and prior can be accessed without any authentication information such as the session cookie. An attacker can send requests to sensitive endpoints as an unauthenticated user to perform critical actions or modify critical configuration parameters.
EPSS: 0.009 (58.7th percentile) — read the EPSS interpretation.
CVSS v3 metric
CVSS v3 base score 7.3 (High). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L.
Affected products
- Fresenius Kabi Agilia Link+ — versions unspecified
- Fresenius-kabi Agilia_connect
- Fresenius-kabi Agilia_connect_firmware
- Fresenius-kabi Agilia_partner_maintenance_software
- Fresenius-kabi Link\+_agilia
- Fresenius-kabi Link\+_agilia_firmware — versions 3.0
- Fresenius-kabi Vigilant_centerium — versions 1.0
- Fresenius-kabi Vigilant_insight — versions 1.0
- Fresenius-kabi Vigilant_mastermed — versions 1.0
Weakness classification (CWE)
References
- ics-cert@hq.dhs.gov (US Government Resource, Third Party Advisory, x_refsource_MISC)
Frequently asked questions
- What is CVE-2021-23233?
- CVE-2021-23233 is a high-severity vulnerability in Fresenius Kabi Agilia Link+, classified under Improper Access Control. CVSS score: 7.3/10. Published 2022-01-21.
- How severe is CVE-2021-23233?
- High severity. CVSS v3 base score is 7.3 out of 10.