Auth bypass in Gehealthcare Apexpro Telemetry Server Firmware
CVE-2020-6964
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X and CARESCAPE Central Station (CSCS) Versions 2.X, the integrated service for keyboard switching of the affected devices could allow attackers to obtain remote keyboard input access without authentication over the network.
Vulnerability class: Broken Authentication
Published · last modified .
CVSS v3 metric
CVSS v3 base score 8.6 (High). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:N.
EPSS exploit prediction
EPSS 0.014, 72.2th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Gehealthcare Apexpro Telemetry Server Firmware
- Gehealthcare Carescape Central Station Mai700 Firmware — versions 1.0, 2.0
- Gehealthcare Carescape Central Station Mas700 Firmware — versions 1.0, 2.0
- Gehealthcare Carescape Telemetry Server Mp100r Firmware
- Gehealthcare Clinical Information Center Mp100d Firmware — versions 4.0, 5.0
- Gehealthcare Clinical Information Center Mp100r Firmware — versions 4.0, 5.0
- Ge Carescape Telemetry Server,apexpro Server,carescape Central Station,clinical Information Center Systems,carescape B450,b650,b850 Monitors — versions ApexPro Telemetry Server,v4.2 & prior,CARESCAPE Telemetry Server, v4.2 & prior,Clinical Information Center,v4.X& 5.X,CARESCAPE Telemetry Server,v4.3,CARESCAPE Central Station,v1.X,CARESCAPE Central Station,v2.X,B450,v2.X,B650,v1.X,B650,v2.X,B850,v1.X,B850,v2.X
Weakness classification (CWE)
References
- www.us-cert.gov/ics/advisories/icsma-20-023-01 (US Government Resource, Third Party Advisory)
- www3.gehealthcare.com/~/media/downloads/us/support/site-planning/site-readiness… (Product)