Vulnerability in Philips Patient Information Center Ix

CVE-2020-16220

In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain syntax) but it does not validate or incorrectly validates that the input complies with the syntax, causing the certificate enrollment service to crash. It does not impact monitoring but prevents new devices from enrolling.

EPSS: 0.003 (20.3th percentile) — read the EPSS interpretation.

CVSS v3 metric

CVSS v3 base score 4.3 (Medium). Vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L.

Affected products

Weakness classification (CWE)

Public proof-of-concept exploits

References

Frequently asked questions

What is CVE-2020-16220?
CVE-2020-16220 is a medium-severity vulnerability in Philips Patient Information Center Ix, classified under CWE-1286. CVSS score: 4.3/10. Published 2020-09-11.
How severe is CVE-2020-16220?
Medium severity. CVSS v3 base score is 4.3 out of 10.
Is CVE-2020-16220 known to be exploited?
1 public proof-of-concept repositories are indexed. Not currently listed in the CISA KEV catalog.