Privilege escalation in Siemens Dca Vantage Analyzer
CVE-2020-15797
- CVSS v3
- 6.8 Medium
- Public exploits
- None known
- EPSS
- 0.004 · 33.3th percentile
- Published
- 2020-10-13
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Improper Access Control could allow an unauthenticated attacker to escape from the restricted environment (“kiosk mode”) and access the underlying operating system. Successful exploitation requires direct physical access to the system.
Vulnerability class: Privilege Escalation
Published · last modified .
CVSS v3 metric
CVSS v3 base score 6.8 (Medium). Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
- Attack Vector
- Physical
AV:P - Attack Complexity
- Low
AC:L - Privileges Required
- None
PR:N - User Interaction
- None
UI:N - Scope
- Unchanged
S:U - Confidentiality
- High
C:H - Integrity
- High
I:H - Availability
- High
A:H
EPSS exploit prediction
EPSS 0.004, 33.3th percentile (last changed ); unchanged over 30 days. About EPSS.
Affected products
- Siemens Dca Vantage Analyzer — versions All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797
- Siemens Dca Vantage Analyzer Firmware
Weakness classification (CWE)
References
- www.siemens-healthineers.com/support-documentation/security-advisory (Vendor Advisory)