Vulnerability in Baxter Phoenix X36
CVE-2020-12048
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 7.5 (High). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.
EPSS exploit prediction
EPSS: 0.005 (36.8th percentile), scored .
Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 37th percentile — 36.8% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.
EPSS over last 30 days · oldest: 0.005 · newest: 0.005 · change: 0.000
Affected products
- Baxter Phoenix X36
- Baxter Phoenix X36 Firmware — versions 3.36, 3.40
- Baxter Phoenix Hemodialysis Delivery System — versions Phoenix Hemodialysis Delivery System SW 3.36 and 3.40
Weakness classification (CWE)
References
- www.us-cert.gov/ics/advisories/icsma-20-170-03 (US Government Resource, Third Party Advisory)
Frequently asked questions
- What is CVE-2020-12048?
- CVE-2020-12048 is a high-severity vulnerability in Baxter Phoenix X36, classified under Cleartext Transmission of Sensitive Information. CVSS score: 7.5/10. Published 2020-06-29.
- How severe is CVE-2020-12048?
- High severity. CVSS v3 base score is 7.5 out of 10.