Vulnerability in Baxter Prismaflex

CVE-2020-12036

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

EPSS: 0.005 (39.7th percentile) — read the EPSS interpretation.

CVSS v3 metric

CVSS v3 base score 7.5 (High). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.

Affected products

Weakness classification (CWE)

References

Frequently asked questions

What is CVE-2020-12036?
CVE-2020-12036 is a high-severity vulnerability in Baxter Prismaflex, classified under Cleartext Transmission of Sensitive Information. CVSS score: 7.5/10. Published 2020-06-29.
How severe is CVE-2020-12036?
High severity. CVSS v3 base score is 7.5 out of 10.