Vulnerability in Baxter Em1200 Firmware
CVE-2020-12008
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. This could allow an attacker with network access to view sensitive data including PHI.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 7.5 (High). Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.
EPSS exploit prediction
EPSS 0.010, 62.5th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Baxter Em1200 Firmware — versions 1.1, 1.2
- Baxter Em2400 Firmware — versions 1.10, 1.11
- Baxter Exactamix Em 2400 & 1200 — versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
Weakness classification (CWE)
References
- www.us-cert.gov/ics/advisories/icsma-20-170-01 (US Government Resource, Third Party Advisory)