Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 7.6 (High). Vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H.

EPSS exploit prediction

EPSS: 0.002 (8.4th percentile), scored .

Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 8th percentile — 8.4% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.

EPSS trend (30 days)EPSS over the last 30 days for CVE-2019-25722: held from 0.002 to 0.002.

EPSS over last 30 days · oldest: 0.002 · newest: 0.002 · change: 0.000