Vulnerability in Philips Intellispace Perinatal
CVE-2019-13546
In IntelliSpace Perinatal, Versions K and prior, a vulnerability within the IntelliSpace Perinatal application environment could enable an unauthorized attacker with physical access to a locked application screen, or an authorized remote desktop session host application user to break-out from the containment of the application and access unauthorized resources from the Windows operating system as the limited-access Windows user. Due to potential Windows vulnerabilities, it may be possible for additional attack methods to be used to escalate privileges on the operating system.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 6.8 (Medium). Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS 0.004, 35.2th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Philips Intellispace Perinatal
- Intellispace Perinatal — versions Versions K and prior
Weakness classification (CWE)
References
- www.us-cert.gov/ics/advisories/icsma-19-297-01 (US Government Resource, Third Party Advisory)