Arbitrary file upload in Bd Alaris Cc Syringe Pump Firmware
CVE-2019-10959
- CVSS v3
- 10.0 Critical
- Public exploits
- None known
- EPSS
- 0.025 · 85.3th percentile
- Published
- 2019-06-13
BD Alaris Gateway Workstation Versions, 1.1.3 Build 10, 1.1.3 MR Build 11, 1.2 Build 15, 1.3.0 Build 14, 1.3.1 Build 13, This does not impact the latest firmware Versions 1.3.2 and 1.6.1, Additionally, the following products using software Version 2.3.6 and below, Alaris GS, Alaris GH, Alaris CC, Alaris TIVA, The application does not restrict the upload of malicious files during a firmware update.
Vulnerability class: Unrestricted File Upload
Published · last modified .
CVSS v3 metric
CVSS v3 base score 10.0 (Critical). Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H.
- Attack Vector
- Network
AV:N - Attack Complexity
- Low
AC:L - Privileges Required
- None
PR:N - User Interaction
- None
UI:N - Scope
- Changed
S:C - Confidentiality
- High
C:H - Integrity
- High
I:H - Availability
- High
A:H
EPSS exploit prediction
EPSS 0.025, 85.3th percentile (last changed ); unchanged over 30 days. About EPSS.
Affected products
- Bd Alaris Cc Syringe Pump Firmware
- Bd Alaris Gateway Workstation Firmware — versions 1.1.3, 1.2, 1.3.0
- Bd Alaris Gh Syringe Pump Firmware
- Bd Alaris Gs Syringe Pump Firmware
- Bd Alaris Tiva Syringe Pump Firmware
- Bd Alaris Gateway Workstation — versions Versions, 1.1.3 Build 10, 1.1.3 MR Build 11, 1.2 Build 15, 1.3.0 Build 14, 1.3.1 Build 13, This does not impact the latest firmware Versions 1.3.2 and 1.6.1, Additionally, the following products using software Version 2.3.6 and below, Alaris GS, Alaris GH, Alaris CC, Alaris TIVA
Weakness classification (CWE)
References
- ics-cert.us-cert.gov/advisories/ICSMA-19-164-01 (US Government Resource, Third Party Advisory, Mitigation)
- www.securityfocus.com/bid/108765 (Third Party Advisory, VDB Entry, vdb-entry)
- www.bd.com/en-us/support/product-security-and-privacy/product-security-bulletin… (Vendor Advisory)