Vulnerability in Medtronic 24950 Mycarelink Monitor
CVE-2018-8868
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 6.2 (Medium). Vector: CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L.
EPSS exploit prediction
EPSS 0.003, 27.4th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Medtronic 24950 Mycarelink Monitor — versions All versions
- Medtronic 24950 Mycarelink Monitor Firmware
- Medtronic 24952 Mycarelink Monitor — versions All versions
- Medtronic 24952 Mycarelink Monitor Firmware
Weakness classification (CWE)
References
- global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-6-28-…
- ics-cert.us-cert.gov/advisories/ICSMA-18-179-01 (US Government Resource, Third Party Advisory)