Vulnerability in Philips Brilliance Ct Big Bore Firmware
CVE-2018-8861
Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) could enable a limited-access kiosk user or an unauthorized attacker to break-out from the containment of the kiosk environment, attain elevated privileges from the underlying Windows OS, and access unauthorized resources from the operating system.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 8.7 (High). Vector: CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:L.
EPSS exploit prediction
EPSS 0.004, 34.5th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Philips Brilliance Ct Big Bore Firmware
- Philips Brilliance Ct Scanners — versions Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior.
- Philips Brilliance Firmware 64
- Philips Brilliance Ict Firmware
- Philips Brilliance Ict Sp Firmware
Weakness classification (CWE)
References
- www.usa.philips.com/healthcare/about/customer-support/product-security (Vendor Advisory)
- www.securityfocus.com/bid/104088 (Third Party Advisory, VDB Entry, vdb-entry)
- ics-cert.us-cert.gov/advisories/ICSMA-18-123-01 (US Government Resource, Third Party Advisory)