Vulnerability in Philips Brilliance Ct Big Bore Firmware
CVE-2018-8857
Philips Brilliance CT software (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) contains fixed credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. An attacker could compromise these credentials and gain access to the system.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 7.8 (High). Vector: CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS 0.003, 23.5th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Philips Brilliance Ct Big Bore Firmware
- Philips Brilliance Ct Scanners — versions Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior.
- Philips Brilliance Firmware 64
- Philips Brilliance Ict Firmware
- Philips Brilliance Ict Sp Firmware
Weakness classification (CWE)
References
- www.usa.philips.com/healthcare/about/customer-support/product-security (Vendor Advisory)
- www.securityfocus.com/bid/104088 (Third Party Advisory, VDB Entry, vdb-entry)
- ics-cert.us-cert.gov/advisories/ICSMA-18-123-01 (US Government Resource, Third Party Advisory)