Vulnerability in Beaconmedaes Scroll Medical Air Systems Firmware
CVE-2018-7510
In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prior to 4107600010.23, passwords are presented in plaintext in a file that is accessible without authentication.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 9.8 (Critical). Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS 0.014, 72.7th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Beaconmedaes Scroll Medical Air Systems Firmware
- Ics-Cert Beaconmedaes Totalalert Scroll Medical Air Systems Web Application — versions All versions prior to version 4107600010.23
Weakness classification (CWE)
References
- ics-cert.us-cert.gov/advisories/ICSMA-18-144-01 (US Government Resource, Third Party Advisory)