Information disclosure in Draeger Delta Xl Firmware
CVE-2018-19014
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 6.5 (Medium). Vector: CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.
EPSS exploit prediction
EPSS 0.008, 56.5th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Draeger Delta Xl Firmware
- Draeger Infinity Delta Firmware
- Draeger Infinity Explorer C700 Firmware
- Draeger Kappa Firmware
- Ics-Cert Dräger Infinity Delta — versions Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions.
Weakness classification (CWE)
References
- ics-cert.us-cert.gov/advisories/ICSMA-19-022-01 (US Government Resource, Third Party Advisory)
- www.securityfocus.com/bid/106683 (Third Party Advisory, VDB Entry, vdb-entry)