Vulnerability in Medtronic 24950 Mycarelink Monitor
CVE-2018-10626
Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 4.4 (Medium). Vector: CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:N.
EPSS exploit prediction
EPSS: 0.004 (27.8th percentile), scored .
Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 28th percentile — 27.8% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.
EPSS over last 30 days · oldest: 0.004 · newest: 0.004 · change: 0.000
Affected products
- Medtronic 24950 Mycarelink Monitor — versions All versions
- Medtronic 24952 Mycarelink Monitor — versions All versions
- Medtronic Mycarelink 24950 Patient Monitor
- Medtronic Mycarelink 24950 Patient Monitor Firmware
- Medtronic Mycarelink 24952 Patient Monitor
- Medtronic Mycarelink 24952 Patient Monitor Firmware
Weakness classification (CWE)
References
- global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-1…
- af854a3a-2127-422b-91ae-364da2661108 (US Government Resource, Third Party Advisory)
- www.securityfocus.com/bid/105042 (VDB Entry, Third Party Advisory, vdb-entry)
- github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2018/icsma-18-219-01.j…
- www.cisa.gov/news-events/ics-medical-advisories/icsma-18-219-01
Frequently asked questions
- What is CVE-2018-10626?
- CVE-2018-10626 is a medium-severity vulnerability in Medtronic 24950 Mycarelink Monitor, classified under Insufficient Verification of Data Authenticity. CVSS score: 4.4/10. Published 2018-08-10.
- How severe is CVE-2018-10626?
- Medium severity. CVSS v3 base score is 4.4 out of 10.