Vulnerability in Medtronic 24950 Mycarelink Monitor
CVE-2018-10626
Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An
attacker who obtains per-product credentials from the monitor and paired
implantable cardiac device information can potentially upload invalid
data to the Medtronic CareLink network.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 4.4 (Medium). Vector: CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:N.
EPSS exploit prediction
EPSS 0.004, 30.4th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Medtronic 24950 Mycarelink Monitor — versions All versions
- Medtronic 24952 Mycarelink Monitor — versions All versions
- Medtronic Mycarelink 24950 Patient Monitor Firmware
- Medtronic Mycarelink 24952 Patient Monitor Firmware
Weakness classification (CWE)
References
- global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-1…
- ics-cert.us-cert.gov/advisories/ICSMA-18-219-01 (US Government Resource, Third Party Advisory)
- www.securityfocus.com/bid/105042 (VDB Entry, Third Party Advisory, vdb-entry)
- github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2018/icsma-18-219-01.j…
- www.cisa.gov/news-events/ics-medical-advisories/icsma-18-219-01