Buffer overflow in Ics-Cert Intellivue Patient Monitors, Avalon Fetal/maternal Monitors

CVE-2018-10601

IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.

Vulnerability class: Buffer Overflow

Published · last modified .

CVSS v3 metric

CVSS v3 base score 8.2 (High). Vector: CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H.

EPSS exploit prediction

EPSS: 0.004 (29.7th percentile), scored .

Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 30th percentile — 29.7% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.

EPSS trend (30 days)EPSS over the last 30 days for CVE-2018-10601: held from 0.004 to 0.004.

EPSS over last 30 days · oldest: 0.004 · newest: 0.004 · change: +0.000

Affected products

Weakness classification (CWE)

References

Frequently asked questions

What is CVE-2018-10601?
CVE-2018-10601 is a high-severity vulnerability in Ics-Cert Intellivue Patient Monitors, Avalon Fetal/maternal Monitors, classified under Stack-based Buffer Overflow. CVSS score: 8.2/10. Published 2018-06-05.
How severe is CVE-2018-10601?
High severity. CVSS v3 base score is 8.2 out of 10.