Vulnerability in I-Sens Smartlog Diabetes Management Software
CVE-2017-13993
An Uncontrolled Search Path or Element issue was discovered in i-SENS SmartLog Diabetes Management Software, Version 2.4.0 and prior versions. An uncontrolled search path element vulnerability has been identified which could be exploited by placing a specially crafted DLL file in the search path. If the malicious DLL is loaded prior to the valid DLL, an attacker could execute arbitrary code on the system. This vulnerability does not affect the connected blood glucose monitor and would not impact delivery of therapy to the patient.
EPSS: 0.018 (75.7th percentile) — read the EPSS interpretation.
CVSS v3 metric
CVSS v3 base score 7.8 (High). Vector: CVSS:3.0/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H.
Affected products
- I-sens Smartlog_diabetes_management_software
- N/a I-sens, Inc. Smartlog Diabetes Management Software — versions i-SENS, Inc. SmartLog Diabetes Management Software
Weakness classification (CWE)
References
- ics-cert@hq.dhs.gov (US Government Resource, Patch, Third Party Advisory, x_refsource_MISC)
- ics-cert@hq.dhs.gov (VDB Entry, Third Party Advisory, vdb-entry, x_refsource_BID)
Frequently asked questions
- What is CVE-2017-13993?
- CVE-2017-13993 is a high-severity vulnerability in I-Sens Smartlog Diabetes Management Software, classified under Unquoted Search Path or Element. CVSS score: 7.8/10. Published 2017-10-05.
- How severe is CVE-2017-13993?
- High severity. CVSS v3 base score is 7.8 out of 10.