Vulnerability in Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump

CVE-2017-12725

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

EPSS: 0.008 (53.9th percentile) — read the EPSS interpretation.

CVSS v3 metric

CVSS v3 base score 5.6 (Medium). Vector: CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L.

Affected products

Weakness classification (CWE)

References

Frequently asked questions

What is CVE-2017-12725?
CVE-2017-12725 is a medium-severity vulnerability in Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump, classified under Use of Hard-coded Credentials. CVSS score: 5.6/10. Published 2018-02-15.
How severe is CVE-2017-12725?
Medium severity. CVSS v3 base score is 5.6 out of 10.