Vulnerability in Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump
CVE-2017-12724
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 8.1 (High). Vector: CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS: 0.013 (69.8th percentile), scored .
Low probability of exploitation in the next 30 days, but still tracked. 70th percentile — 69.8% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.
EPSS over last 30 days · oldest: 0.013 · newest: 0.013 · change: 0.000
Affected products
- Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump — versions 1.1, 1.5, 1.6
- Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump — versions Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
Weakness classification (CWE)
References
- ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A (US Government Resource, Third Party Advisory)
- www.securityfocus.com/bid/100665 (VDB Entry, Third Party Advisory, vdb-entry)
Frequently asked questions
- What is CVE-2017-12724?
- CVE-2017-12724 is a high-severity vulnerability in Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump, classified under Use of Hard-coded Credentials. CVSS score: 8.1/10. Published 2018-02-15.
- How severe is CVE-2017-12724?
- High severity. CVSS v3 base score is 8.1 out of 10.