Vulnerability in Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump

CVE-2017-12724

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.

EPSS: 0.013 (68.0th percentile) — read the EPSS interpretation.

CVSS v3 metric

CVSS v3 base score 8.1 (High). Vector: CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H.

Affected products

Weakness classification (CWE)

References

Frequently asked questions

What is CVE-2017-12724?
CVE-2017-12724 is a high-severity vulnerability in Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump, classified under Use of Hard-coded Credentials. CVSS score: 8.1/10. Published 2018-02-15.
How severe is CVE-2017-12724?
High severity. CVSS v3 base score is 8.1 out of 10.