Vulnerability in Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump
CVE-2017-12724
- CVSS v3
- 8.1 High
- Public exploits
- None known
- EPSS
- 0.013 · 71.4th percentile
- Published
- 2018-02-15
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 8.1 (High). Vector: CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS 0.013, 71.4th percentile (last changed ); unchanged over 30 days. About EPSS.
Affected products
- Smiths-Medical Medfusion 4000 Wireless Syringe Infusion Pump — versions 1.1, 1.5, 1.6
- Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump — versions Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
Weakness classification (CWE)
References
- ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A (US Government Resource, Third Party Advisory)
- www.securityfocus.com/bid/100665 (Third Party Advisory, VDB Entry, vdb-entry)