Vulnerability in Abbott Accent Firmware
CVE-2017-12716
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 6.5 (Medium). Vector: CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.
EPSS exploit prediction
EPSS 0.002, 15.3th percentile (last changed ); unchanged over 30 days. About EPSS.
Affected products
- Abbott Accent Firmware
- Abbott Accent Mri Firmware
- Abbott Accent St Firmware
- Abbott Anthem Firmware
- Abbott Laboratories Accent And Anthem — versions All versions of pacemakers manufactured prior to August 28, 2017
Weakness classification (CWE)
References
- ics-cert.us-cert.gov/advisories/ICSMA-17-241-01 (US Government Resource, Third Party Advisory)
- www.securityfocus.com/bid/100523 (Third Party Advisory, VDB Entry, vdb-entry)