Vulnerability in Blipcare Wi-Fi Blood Pressure Monitor Firmware

CVE-2017-11580

CVSS v3
6.5 Medium
Public exploits
None known
EPSS
0.014 · 73.5th percentile
Published
2019-07-02

Blipcare Wifi blood pressure monitor BP700 10.1 devices allow memory corruption that results in Denial of Service. When connected to the "Blip" open wireless connection provided by the device, if a large string is sent as a part of the HTTP request in any part of the HTTP headers, the device could become completely unresponsive. Presumably this happens as the memory footprint provided to this device is very small. According to the specs from Rezolt, the Wi-Fi module only has 256k of memory. As a result, an incorrect string copy operation using either memcpy, strcpy, or any of their other variants could result in filling up the memory space allocated to the function executing and this would result in memory corruption. To test the theory, one can modify the demo application provided by the Cypress WICED SDK and introduce an incorrect "memcpy" operation and use the compiled application on the evaluation board provided by Cypress semiconductors with exactly the same Wi-Fi SOC. The results were identical where the device would completely stop responding to any of the ping or web requests.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 6.5 (Medium). Vector: CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H.

Attack Vector
Adjacent AV:A
Attack Complexity
Low AC:L
Privileges Required
None PR:N
User Interaction
None UI:N
Scope
Unchanged S:U
Confidentiality
None C:N
Integrity
None I:N
Availability
High A:H

EPSS exploit prediction

EPSS 0.014, 73.5th percentile (last changed ); unchanged over 30 days. About EPSS.

Affected products

Weakness classification (CWE)

Public exploits and proof-of-concept code

Also mentioned in 1 other repository (ethanhunnt/IoT_vulnerabilities). These reference the CVE in a list, notes or a scanner and are not exploit code for it.

References