Buffer overflow in Ishekar Endoscope Camera
CVE-2017-10723
Recently it was discovered as a part of the research on IoT devices in the most recent firmware for Shekar Endoscope that an attacker connected to the device Wi-Fi SSID can exploit a memory corruption issue and execute remote code on the device. This device acts as an Endoscope camera that allows its users to use it in various industrial systems and settings, car garages, and also in some cases in the medical clinics to get access to areas that are difficult for a human being to reach. Any breach of this system can allow an attacker to get access to video feed and pictures viewed by that user and might allow them to get a foot hold in air gapped networks especially in case of nation critical infrastructure/industries. The firmware contains binary uvc_stream that is the UDP daemon which is responsible for handling all the UDP requests that the device receives. The client application sends a UDP request to change the Wi-Fi name which contains the following format: "SETCMD0001+0001+[2 byte length of wifiname]+[Wifiname]. This request is handled by "control_Dev_thread" function which at address "0x00409AE0" compares the incoming request and determines if the 10th byte is 01 and if it is then it redirects to 0x0040A74C which calls the function "setwifiname". The function "setwifiname" uses a memcpy function but uses the length of the payload obtained by using strlen function as the third parameter which is the number of bytes to copy and this allows an attacker to overflow the function and control the $PC value.
Vulnerability class: Buffer Overflow
EPSS: 0.026 (83.5th percentile) — read the EPSS interpretation.
CVSS v3 metric
CVSS v3 base score 8.8 (High). Vector: CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H.
Affected products
- Ishekar Endoscope_camera
- Ishekar Endoscope_camera_firmware
- N/a — versions n/a
Weakness classification (CWE)
Public proof-of-concept exploits
References
- cve@mitre.org (mailing-list, x_refsource_BUGTRAQ, Mailing List, Third Party Advisory)
- cve@mitre.org (VDB Entry, Third Party Advisory, x_refsource_MISC)
- cve@mitre.org (Exploit, Third Party Advisory, x_refsource_MISC)
Frequently asked questions
- What is CVE-2017-10723?
- CVE-2017-10723 is a high-severity vulnerability in Ishekar Endoscope Camera, classified under Improper Restriction of Operations within the Bounds of a Memory Buffer. CVSS score: 8.8/10. Published 2019-06-17.
- How severe is CVE-2017-10723?
- High severity. CVSS v3 base score is 8.8 out of 10.
- Is CVE-2017-10723 known to be exploited?
- 1 public proof-of-concept repositories are indexed. Not currently listed in the CISA KEV catalog.