Vulnerability in Bd Alaris 8015 Pc Unit

CVE-2016-9355

An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7. An unauthorized user with physical access to an Alaris 8015 PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling an Alaris 8015 PC unit and accessing the device's flash memory. Older software versions of the Alaris 8015 PC unit, Version 9.5 and prior versions, store wireless network authentication credentials and other sensitive technical data on the affected device's removable flash memory. Being able to remove the flash memory from the affected device reduces the risk of detection, allowing an attacker to extract stored data at the attacker's convenience.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 5.3 (Medium). Vector: CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:N/A:N.

EPSS exploit prediction

EPSS 0.005, 44.1th percentile (scored ); unchanged over 30 days. About EPSS.

Affected products

  • Bd Alaris 8015 Pc Unit — versions 9.7
  • Bd Alaris 8015 Insufficiently Protected Credentials Vulnerabilities — versions BD Alaris 8015 Insufficiently Protected Credentials Vulnerabilities

Weakness classification (CWE)

Public exploits and proof-of-concept code

1 public repository with exploit or proof-of-concept code for this CVE:

  • nadafarafat/bd-alaris-firmware-analysis (last pushed 2026-06-10): Firmware security analysis of BD Alaris 8015 infusion pump (CVE-2016-9355). Identified 6 compound vulnerabilities including plaintext Wi-Fi credentials, HTTP firmware updates, and disabled integrity…

References