Vulnerability in Hospira Plum A+3 Infusion System
CVE-2015-3956
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 9.8 (Critical). Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.
EPSS exploit prediction
EPSS 0.010, 61.9th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Hospira Plum A+3 Infusion System — versions <= 13.6
- Hospira Plum A+ Infusion System — versions <= 13.4
- Hospira Symbiq Infusion System — versions <= 3.13
- Pifzer Plum A\+3 Infusion System Firmware
- Pifzer Plum A\+ Infusion System Firmware
- Pifzer Symbiq Infusion System Firmware
Weakness classification (CWE)
References
- ics-cert.us-cert.gov/advisories/ICSA-15-161-01 (US Government Resource, Third Party Advisory, Mitigation)