Auth bypass in Hospira Plum A+3 Infusion System

CVE-2015-3954

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 9.8 (Critical). Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.

EPSS exploit prediction

EPSS: 0.020 (80.3th percentile), scored .

Low probability of exploitation in the next 30 days, but still tracked. 80th percentile — 80.3% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.